Troubleshooting | Immunofluorescence
Follow our immunofluorescence troubleshooting guide to quickly identify and get solutions for common protocol issues.
Identify the problem with your immunofluorescence staining from the options below:
Weak or No Staining
Incorrect light source/filter set:
- Ensure your microscope is equipped with the correct light source and filter set for the fluorophore you have chosen.
Gain/exposure is too low:
- Turn up the gain and/or increase the exposure time to ensure you are capturing any signal present.
Fluorescent tag bleached:
- Avoid over exposure of the slide to light sources for extended periods. Always store slides in the dark.
Cell/tissues are over fixed:
- Reduced the duration of fixation.
- Perform antigen retrieval to unmask the epitope.
Cells were not permeabilized:
- Methanol and acetone fixation will permeabilize cells.
- If using formaldehyde, permeabilize cells with 0.2% Triton X-100.
Tissue/cells dried out:
- Samples must be kept covered in liquid throughout the staining process.
Not enough primary antibody:
- Use a higher concentration of antibody.
- Incubate longer.
The primary and secondary antibodies are incompatible:
- The secondary antibody should be raised against the host of the primary antibody. For example, if the primary antibody is a Mouse Anti-HSP70, use an Anti-Mouse secondary antibody (ie. Goat Anti-Mouse).
- Isotypes should also be compatible.
Suitability of the primary antibody:
- Confirm that the antibody has been validated in IHC, and specifically what type- formalin fixed, paraffin-embedded, fresh frozen, etc.
- Test the antibody in a western blot to make sure it hasn’t been damaged.
Slide storage issues:
- Samples should be imaged shortly after processing as the signal decreases over time. Store slides at 4⁰C in the dark if needed.
Antibody Storage issues:
- Freeze/thaw cycles are detrimental and can cause degradation. It is best to create aliquots of smaller amounts as soon as the product arrives at your location.
- Antibody was not stored as recommended. Unfortunately this might require a new vial to be used instead.
- If the secondary was not stored in the dark (when using immunofluorescence), a new vial will need to be used instead.
The protein is not present in the tissues being tested:
- Run a positive control.
- If the protein is present, but not abundantly, use an amplification step to maximize the signal.
- Increase the duration of incubation of the primary antibody with the sample.
High Background
Autofluorescence:
- Check to see if there is any fluorescence in an unstained section of the processed tissue. If there is, then this is autofluorescence in the tissue.
- Avoid glutaraldehyde fixative or wash with 0.1% sodium borohydride in PBS to remove free aldehyde groups.
- May be due to endogenous molecules (FAD, FMN, NADH, lipofuscin), and will require pre-photobleaching, treatment with sudan black, or cupric sulfate.
Tissue is too thick:
- Consider using thinner tissue sections.
Antibody Concentration is too high:
- Reduce the concentration of the primary and/or secondary antibody used.
Secondary is binding non-specifically:
- Run a secondary control without the primary. If there is staining, then change the secondary.
Blocking is insufficient:
- Increase the blocking incubation period and consider changing the blocking agent.
Amplification:
- Reduce amplification incubation time and dilute the secondary antibody.
Insufficient washing:
- Proper washing of the tissue between steps is critical. Ensure you are following the protocol guidelines for wash steps.
Non-specific Staining
Spectral overlap:
- If imaging more than one fluorescent probe, the fluorophores may have excitation and emission spectra that overlap. Adjust your light sources and filters to pick up only one signal at a time. If this is not possible, choose new fluorophores that do not have spectral overlap.
Antibody Concentrations too high:
- Try reducing the concentration, and the incubation period.
The primary is raised again the same species as the tissues stained (eg. Mouse on mouse):
- Try using a primary that is raised against a different species. Otherwise try to block the endogenous IgG with serum from the same species as the secondary. You can also try to incubate sections with 1% Triton to clean the tissues. Or Use TBS-Tween 20 as a washing buffer, rather than using PBS-Tween 20.
Aggregates:
- Spin down secondary antibodies in a microcentrifuge to move aggregates to the bottom of the tube. Take from the top.